Hardware - HWA2023-06

Informations sur le produit

Info about measurement PCs (ibaRackline / ibaDeskline)

Fond vert avec design d’interface numérique circulaire.

To ensure trouble-free measurement operation, the status of the “C-States” in the BIOS of the measurement PC must be set to “Disabled”.

For ibaRackline and ibaDeskline systems delivered by iba AG from Q2/2023, this is already set by default in the BIOS.

To avoid unstable behavior, implausible measured values and FOB I/O cards that are no longer recognized, please check the status of the “C-States” in the BIOS and set it to “Disabled” if necessary.

Bios Settings:

→ Advanced Menu

→ CPU Configuration

→ CPU - Power Management Control

→ C states ->Set to “Disabled”

Be sure to perform a cold start after the change.